Scintillation Survey MeterFEATURES

• MEASUREMENT MODE, DOSE RATE, COUNT RATE, DOSE

• MEASUREABLE FOR PULSE DOSE SUCH AS X-RAY PHOTOGRAPHY

• SINGLE CHANNEL PEAK ANALYSZNG FUNCTION (2ch) FOR EACH NUCLIDE

• MEASURED VALUES ARE DISPLAYED

• COUNT MONITOR WITH SOUND BUZZER

• COUNT MONITOR LIGHT WITH LED

 

The RAD-SUR-SCINT is a scintillation survey meter (energy compensated) to measure gamma (X) rays.

SPECIFICATIONS

Radiation detected : Gamma (X) rays

Detector : NaI (Tl) scintillator φ25.4xφ25.4mm / f1 x t 1 in.: Measurement range

Dose rate : BG to 75.000 μSv/h ref. 137Cs

Dose : 0.000 to 99.999 μSv

Counting rate : 0 to 100000 s-1

SCA1 count rate : 0 to 100000 s-1

SCA2 count rate : 0 to 100000 s-1

Accuracy : ±20 %

Energy range : 50 to 3000 keV

Energy response : ±25% (50 to 3000 keV)

Angular dependence : ±20 % (-90 to 90°)

Response time Dose rate : 4 sec (1 to 10 μSv/h)

Counting rate : 20 sec (1 to 10 s-1)

Display : Organic EL display

Historical data storage : 1500 points approx.

Battery life : Displays continuously 10 hours or more

Displays intermittently 20 hours or more (displays 1min, during 5min.)

Data output : USB interface for PC

Power supply : AA Alkaline batteries (LR6) Å~6 cells

100 to 240V AC with AC adaptor

AA Ni-H rechargeable battery Å~6 cells (option)

Operating temperature : 32 to 104°F / 0 to 40°C

Operating humidity : 90 % (non-condensing)