FEATURES
• MEASUREMENT MODE, DOSE RATE, COUNT RATE, DOSE
• MEASUREABLE FOR PULSE DOSE SUCH AS X-RAY PHOTOGRAPHY
• SINGLE CHANNEL PEAK ANALYSZNG FUNCTION (2ch) FOR EACH NUCLIDE
• MEASURED VALUES ARE DISPLAYED
• COUNT MONITOR WITH SOUND BUZZER
• COUNT MONITOR LIGHT WITH LED
The RAD-SUR-SCINT is a scintillation survey meter (energy compensated) to measure gamma (X) rays.
SPECIFICATIONS
Radiation detected : Gamma (X) rays
Detector : NaI (Tl) scintillator φ25.4xφ25.4mm / f1 x t 1 in.: Measurement range
Dose rate : BG to 75.000 μSv/h ref. 137Cs
Dose : 0.000 to 99.999 μSv
Counting rate : 0 to 100000 s-1
SCA1 count rate : 0 to 100000 s-1
SCA2 count rate : 0 to 100000 s-1
Accuracy : ≤ ±20 %
Energy range : 50 to 3000 keV
Energy response : ±25% (50 to 3000 keV)
Angular dependence : ≤ ±20 % (-90 to 90°)
Response time Dose rate : 4 sec (1 to 10 μSv/h)
Counting rate : 20 sec (1 to 10 s-1)
Display : Organic EL display
Historical data storage : 1500 points approx.
Battery life : Displays continuously 10 hours or more
Displays intermittently 20 hours or more (displays 1min, during 5min.)
Data output : USB interface for PC
Power supply : AA Alkaline batteries (LR6) Å~6 cells
100 to 240V AC with AC adaptor
AA Ni-H rechargeable battery Å~6 cells (option)
Operating temperature : 32 to 104°F / 0 to 40°C
Operating humidity : ≤ 90 % (non-condensing)